Summary
Type: TLC
Application #: 401460
Catalog#: 1715026 - Silica G HR TLC Plates, glass backed, 250um, 20x20cm, 25/pk
Detection: UV 254 nm
Rf values: cortexone 0.69, cortexolone 0.50, corticosterone 0.45, cortisone 0.41, cortisol 0.27UV 254 nm
Conditions: Eluent: chloroform / ethanol / water (87:13:1, v/v)
Migration distance: 15 cm in 45 min, ascending
Authors: Hartmann F
Source: Res exp Med 161 (1973) 165 and F Hartmann F Jentze
Details
Legend: Rf values: cortexone 0.69, cortexolone 0.50, corticosterone 0.45, cortisone
0.41, cortisol 0.27